Flame emission and atomic absorption – line spectra and calibration
ChemistryInorganic ChemistryAges 17–18Virtual lab
How do you mainly use this simulation?
Use it with your class: students open it from a link or a QR code, with no account needed.
Spray metal salt solutions into a flame and read their emission line spectra through a spectroscope to identify the metals in a mixture, even when one flame colour hides another. Then use an atomic absorption spectrometer with a hollow-cathode lamp of the analyte metal: measure standards, build a calibration line inside the working range, and find lead in waste water and sodium in a sports drink. Lines read and absorbances go into tables.
Lesson: Flame emission spectroscopy and atomic absorption spectroscopy (AAS)
What it shows
Every metal gives out light of its own wavelengths when its atoms are heated in a flame, because electrons fall back between fixed energy levels. A spectroscope spreads this light into a line spectrum, and comparing the lines with reference spectra identifies the metals in a mixture, which flame colours alone cannot do. In atomic absorption the same atoms, still in their ground state, absorb light from a lamp made of that metal. Absorbance is proportional to concentration only inside a working range, so standards, a calibration line and sometimes a dilution are needed. The line data are real; brightness and noise are modelled.
How to use
Flame emission tab: choose a Sample and tap the lines of its spectrum to record their wavelengths, then compare them with the reference spectra. Atomic absorption tab: choose the Lamp and a Solution, press Aspirate and read, and build the calibration graph from the blank and the standards. For a sample, choose a Dilution that brings its reading inside the working range before you calculate its concentration.
Parameters you can change
- View Flame emission, Atomic absorption
- Emission sample Lithium chloride, LiCl, Sodium chloride, NaCl, Potassium chloride, KCl, Calcium chloride, CaCl₂, Strontium chloride, SrCl₂, Barium chloride, BaCl₂, Copper(II) chloride, CuCl₂, Mixture of NaCl and KCl, Unknown X
- AAS lamp Lead (Pb), Copper (Cu), Sodium (Na)
- AAS solution Blank, Standard 1, Standard 2, Standard 3, Standard 4, Standard 5, Standard 6, Standard 7, Industrial waste water, Sports drink
- Dilution ×1, ×2, ×5, ×10, ×20, ×50, ×100, ×200, ×500, ×1000
Questions to explore
- Why can a spectroscope find potassium in a mixture whose flame looks only yellow?
- Why must the AAS lamp be made of the same metal as the one being measured?
- Why is a sample diluted until its absorbance falls between the standards?